The Zimmer ATS 750 Automatic Tourniquet System (#60-0750-101-00) includes tubing, a new internal battery, and power cord.
Warranty: 90 Days - Parts and Labor.
Zimmer Tourniquet systems offer built-in proprietary safety features as recommended by the association of periOperative Registered Nurses.
- Self-check calibration automatically checks the accuracy of the machine calibration each time the unit is powered on.
- Cuff alert notifies user of cuff status when attempt is made to set the machine to stand-by.
- Cuff lockout feature prompts the user to confirm the deflation command during bilateral or IVRA procedure in order to reduce complications due to sudden cuff deflation of the second cuff.
Features
Zimmer ATS 750 Automatic Tourniquet System Features
- Uses ambient air.
- Uses electronic and microprocessor regulated pressure.
-
Single Port System: Uses single port, single bladder cuff.
- 30 minute battery back-up provides mobility by helping ensure continuous power.
- Visual and audible alarms activated when an unsafe condition is detected.
- Self-check calibration each time the A.T.S. 750 System is switched on.
Specifications
Zimmer ATS 750 Automatic Tourniquet System Specifications
-
Dimensions: 7.9" (D) x 7" (W) x 8" (H) (20.1 cm x 17.8 cm x 20.3 cm).
-
Weight: 7.7 lb (3.5 kg).
- Timer:
- Time Alarm Set Ranges: 5 to 240 minutes; 5 minutes increments.
- Timer Accuracy: 0.25% of elapsed time.
- Power:
- Line Voltage Range: 100 to 240 VAC, 50/60 Hz, Auto Switching.
- Line Current: 670 mA RMS @ 120 VAC.
- Input Power: 53 Watts Typical.
- Power Cord: Type SJT, AWG 16, 14 ft (4.27 m).
- Power Plug: Hospital grade, 3 prong straight blade, 15 Amp.
- Pressure:
- Cuff Pressure Range: 50 to 475 mmHg, 5 mmHg increments.
- Pressure Accuracy: +/- 5 mmHg (50 to 475 mmHg).
- Pressure Regulation: +/- 6 mmHg of set point (10 second average under non-transient conditions without external leaks).
- Maximum Pressures: 475 mmHg cuffs.
- Battery:
- Battery Type: Rechargeable, 12 VDC sealed lead acid, 1.2 Amp hours.
- Battery Discharge Time: Unit will operate on battery power for 30 minutes minimum with fully charge batteries.
- Battery Recharge Time: 24 hours (Maximum) Unit should be plugged in 24 hours before initial use.
Literature
Zimmer ATS 750 Automatic Tourniquet System Literature
Zimmer A.T.S. Automatic Tourniquet System Brochure