Carefusion Alaris BD 8110 Specifications
Bolus Volume, Maximum after Occlusion
Without Pressure Sensing Disc
Pressure Setting Low: 0.512 Bolus Volume (mL)
Pressure Setting Medium: 0.776 Bolus Volume (mL)
Pressure Setting High: 1.103 Bolus Volume (mL)
With Pressure Sensing Disc
Pressure Setting 300 mmHg: Back Off Disabled: 0.462; Back Off Enabled: 0.126 Pressure Setting 500 mmHg: Back Off Disabled: 0.575; Back Off Enabled: 0.331 Pressure Setting 1000 mmHg: Back Off Disabled: 0.839; Back Off Enabled: 0.323
Critical Volume
Maximum over-infusion which can occur in the event of a single-fault condition will not exceed 2% of nominal syringe fill volume during loading and 1% of maximum syringe travel after syringe loading.